This case report describes a quick and safe method to successfully retrieve an undeflatable angioplasty balloon via the transradial access site. The article also presents potential mechanisms of contrast media entrapment and bench tests of guidewire-assisted balloon puncture. After successful stent implantation in the proximal right coronary artery in a 74-year-old female patient referred for acute coronary syndrome, the balloon catheter became undeflatable for an unknown reason. Several attempts to burst the balloon in the guide catheter and the aorta failed. After a pullback into the radial artery, the still inflated balloon became stuck again and was unable to be retrieved through the sheath. Finally, transcutaneous puncture and aspiration of contrast media with a 23 gauge syringe needle through the skin in the right forearm allowed deflation and successful removal the balloon.
机译:该病例报告描述了一种快速安全的方法,可以通过the骨入路部位成功取回无法放气的血管成形术球囊。文章还介绍了造影剂截留的潜在机制以及导丝辅助球囊穿刺的台架试验。在一名因急性冠状动脉综合征而病的74岁女性患者中,在成功地将支架植入右冠状动脉近端后,由于未知原因,球囊导管无法放气。几次尝试使导管在引导导管和主动脉中破裂均失败。撤回the动脉后,仍然膨胀的球囊再次卡住,无法通过护套取回。最后,使用23号注射器针头经皮穿刺并通过右前臂皮肤抽吸造影剂,可以放气并成功取出球囊。