Direct Measurement of Mammographic X-Ray Spectra with a Digital CdTe Detection System

机译:用数字CdTe检测系统直接测量乳腺X射线光谱

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In this work we present a detection system, based on a CdTe detector and an innovative digital pulse processing (DPP) system, for high-rate X-ray spectroscopy in mammography (1–30 keV). The DPP system performs a height and shape analysis of the detector pulses, sampled and digitized by a 14-bit, 100 MHz ADC. We show the results of the characterization of the detection system both at low and high photon counting rates by using monoenergetic X-ray sources and a nonclinical X-ray tube. The detection system exhibits excellent performance up to 830 kcps with an energy resolution of 4.5% FWHM at 22.1 keV. Direct measurements of clinical molybdenum X-ray spectra were carried out by using a pinhole collimator and a custom alignment device. A comparison with the attenuation curves and the half value layer values, obtained from the measured and simulated spectra, from an ionization chamber and from a solid state dosimeter, also shows the accuracy of the measurements. These results make the proposed detection system a very attractive tool for both laboratory research, calibration of dosimeters and advanced quality controls in mammography.
机译:在这项工作中,我们提出了一种基于CdTe检测器和创新的数字脉冲处理(DPP)系统的检测系统,用于乳腺X线照相术(1–30 keV)的高速率X射线光谱。 DPP系统对检测器脉冲进行高度和形状分析,并由14位100 MHz ADC进行采样和数字化。我们通过使用单能X射线源和非临床X射线管显示了在低光子计数率和高光子计数率下的检测系统的表征结果。该检测系统在22.1 keV的能量分辨率为4.5%FWHM时,表现出高达830 kcps的出色性能。通过使用针孔准直仪和定制的对准设备对临床钼X射线光谱进行直接测量。从测量和模拟光谱,电离室和固态剂量计获得的衰减曲线和半值层值的比较也显示了测量的准确性。这些结果使所提出的检测系统成为用于实验室研究,剂量计校准和乳房X线照相术中先进质量控制的极具吸引力的工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号