摘要

Matrix effects on the microcystin-LR fluorescent immunoassay based on the evanescent wave all-fiber immunosensor (EWAI) and their elimination methods were studied. The results indicated that PBS and humic acid did not affect the monitoring of samples under the investigated conditions. When the pH was less than 6 or higher than 8, the fluorescence signals detected by immunosensor systems were obviously reduced with the decrease or increase of pH. When the pH ranged from 6 to 8, IC50 and the linear working range of MC-LR calculated from the detection curves were 1.01∼1.04 μg/L and 0.12∼10.5 μg/L, respectively, which was favourable for an MC-LR immunoassay. Low concentrations of Cu2+ rarely affected the detection performance of MC-LR. When the concentration of CuSO4 was higher than 5 mg/L, the fluorescence signal detected by EWAI clearly decreased, and when the concentration of CuSO4 was 10 mg/L, the fluorescence signal detected was reduced by 70%. The influence of Cu2+ on the immunoassay could effectively be compromised when chelating reagent EDTA was added to the pre-reaction mixture.
机译:研究了基于e逝波全纤维免疫传感器(EWAI)的基质对微囊藻毒素-LR荧光免疫分析的影响及其消除方法。结果表明,在所研究的条件下,PBS和腐殖酸不会影响样品的监测。当pH值小于6或大于8时,随着pH值的降低或升高,免疫传感器系统检测到的荧光信号明显降低。当pH为6〜8时,IC50和根据检测曲线计算的MC-LR线性工作范围分别为1.01〜1.04μg/ L和0.12〜10.5μg/ L,有利于MC-LR免疫分析。低浓度的Cu 2 + 很少影响MC-LR的检测性能。当CuSO4的浓度高于5 mg / L时,EWAI检测到的荧光信号明显降低,而当CuSO4的浓度为10 mg / L时,检测到的荧光信号降低70%。当螯合剂EDTA加入到反应前混合物中时,Cu 2 + 对免疫分析的影响可以有效地减弱。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号