Determination of Antimony (III) in Real Samples by Anodic Stripping Voltammetry Using a Mercury Film Screen-Printed Electrode

机译:汞膜丝网印刷电极阳极溶出伏安法测定真实样品中的锑(III)

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This paper describes a procedure for the determination of antimony (III) by differential pulse anodic stripping voltammetry using a mercury film screen-printed electrode as the working electrode. The procedure has been optimized using experimental design methodology. Under these conditions, in terms of Residual Standard Deviation (RSD), the repeatability (3.81 %) and the reproducibility (5.07 %) of the constructed electrodes were both analyzed. The detection limit for Sb (III) was calculated at a value of 1.27×10−8 M. The linear range obtained was between 0.99 × 10−8 − 8.26 × 10−8 M. An analysis of possible effects due to the presence of foreign ions in the solution was performed and the procedure was successfully applied to the determination of antimony levels in pharmaceutical preparations and sea water samples.
机译:本文介绍了一种使用汞薄膜丝网印刷电极作为工作电极的差分脉冲阳极溶出伏安法测定锑(Ⅲ)的方法。该程序已使用实验设计方法进行了优化。在这些条件下,就残留标准偏差(RSD)而言,分析了所构建电极的重复性(3.81%)和再现性(5.07%)。计算得出Sb(III)的检出限为1.27×10 -8 M。线性范围为0.99×10 -8 -8.26×10 -8 M.对溶液中存在外来离子可能产生的影响进行了分析,并将该方法成功地用于测定药物制剂和海水样品中的锑含量。