摘要

Purpose. To evaluate the influence of effective energy on computed tomography (CT) number in monoenergetic images (MEIs). Methods. Three bottle phantoms filled with water, oil, and a contrast agent were scanned at 100 and 140 kVp tube energy with a dual-source CT scanner. Cardiac dual-energy CT data was collected from 17 patients. CT numbers were measured in the 3 phantom materials and in the left ventricular cavity, myocardium, pericardial fat, and vertebral bone in MEIs from 40 to 190 keV. Results. In the phantoms, the mean CT number increased in oil whereas it decreased in the contrast agent as the energy level increased (P < 0.001). In clinical subjects, the mean CT numbers for the left ventricular cavity, myocardium, and vertebral bone were highest in the 40 keV images (P < 0.001) and decreased as the energy level increased. In contrast, the CT number for pericardial fat was lowest in the 40 keV images (P < 0.001) and increased with increasing energy. Conclusions. The influence of effective energy on CT number varies with material and tissue type in monoenergetic cardiac imaging, which could evaluate tissue characteristics through assessment of the changes in CT number associated with effective energies.
机译:目的。评估有效能量对单能图像(MEI)中计算机断层扫描(CT)数量的影响。方法。用双源CT扫描仪分别以100和140 kVp的管能量扫描三个装满水,油和造影剂的瓶体模。从17例患者中收集了心脏双能CT数据。在40到190 keV的MEI中,测量了3种幻象材料以及左心室腔,心肌,心包脂肪和椎骨的CT数。结果。在体模中,油中的平均CT数增加,而造影剂中的平均CT数随着能量水平的增加而降低(P <0.001)。在临床受试者中,在40 cavitykeV图像中,左心室,心肌和椎骨的平均CT数最高(P <0.001),并且随着能量水平的增加而降低。相反,心包脂肪的CT数在40 inkeV图像中最低(P <0.001),并且随着能量的增加而增加。结论。在单能心脏成像中,有效能量对CT数的影响随材料和组织类型的不同而变化,这可以通过评估与有效能量相关的CT数的变化来评估组织特征。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号