Determine the Dose Distribution Using Ultrasound Parameters in MAGIC-f Polymer Gels

机译:使用超声参数确定MAGIC-f聚合物凝胶中的剂量分布

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In this study, using methacrylic and ascorbic acid in gelatin initiated by copper (MAGIC-f) polymer gel after megavoltage energy exposure, the sensitivity of the ultrasound velocity and attenuation coefficient dose-dependent parameters was evaluated. The MAGIC-f polymer gel was irradiated under 1.25 MeV cobalt-60, ranging from 0 to 60 Gy in 2-Gy steps, and received dose uniformity and accuracy of ±2%. After calibration of the ultrasonic systems with a frequency of 500 kHz, the parameters of ultrasound velocity and attenuation coefficient of the irradiated gel samples were measured. According to the dose–response curve, the ability of ultrasonic parameters was evaluated in dose rate readings. Based on a 4-order polynomial curve, fitted on the dose–response parameters of ultrasound velocity and attenuation coefficient and observed at 24 hours after irradiation, ultrasonic parameters had more sensitivity. The sensitivity of the dose–velocity and dose-attenuation coefficient curves was observed as 50 m/s/Gy and 0.06 dB/MHz/Gy over the linear range of 4 to 44 Gy, respectively. The ultrasonic parameters at 5°C, 15°C, and 25°C on the gel dosimeter after 0 to 60 Gy irradiation showed that readings at 25°C have higher sensitivity compared to 15°C and 5°C. Maximum sensitivity time and temperature readings of the MAGIC-f ultrasonic parameters were concluded 24 hours after irradiation and at a temperature of 25°C.
机译:在这项研究中,使用甲基丙烯酸和抗坏血酸在铜(MAGIC-f)聚合物凝胶在高电压能量暴露后引发的明胶中,评估了超声速度的灵敏度和衰减系数的剂量依赖性参数。在1.25 MeV钴60下,以2 Gy的步长在0至60 Gy的范围内辐照MAGIC-f聚合物凝胶,剂量均匀度和精度为±2%。在以500 kHz的频率对超声系统进行校准之后,测量辐照的凝胶样品的超声速度和衰减系数的参数。根据剂量反应曲线,在剂量率读数中评估超声参数的能力。根据4阶多项式曲线,拟合超声速度和衰减系数的剂量响应参数,并在照射后24小时观察到,超声参数具有更高的灵敏度。在4至44 Gy的线性范围内,剂量-速度和剂量衰减系数曲线的灵敏度分别为50 m / s / Gy和0.06 dB / MHz / Gy。 0至60 Gy辐射后,凝胶剂量计在5°C,15°C和25°C的超声参数表明,与15°C和5°C相比,在25°C的读数具有更高的灵敏度。辐照后24小时和25°C下得出MAGIC-f超声参数的最大灵敏度时间和最大温度读数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号