Inductive powering of implantable medical devices involves numerous factors acting on the system efficiency and safety in adversarial ways. This paper lightens up their role and identifies a procedure enabling the system design. The latter enables the problem to be decoupled into four principal steps: the frequency choice, the magnetic link optimization, the secondary circuit and then finally the primary circuit designs. The methodology has been tested for the powering system of a device requirering a power of 300mW and implanted at a distance of 15 to 30mm from the outside power source. It allowed the identification of the most critical parameters. A satisfying efficiency of 34% was reached at 21mm and tend to validate the proposed design procedure.
机译:植入式医疗设备的感应供电涉及许多因素,它们以对抗的方式作用于系统效率和安全性。本文减轻了它们的作用,并确定了启用系统设计的过程。后者使问题可以分解为四个主要步骤:频率选择,磁链优化,次级电路,最后是初级电路设计。该方法已针对要求功率为300mW的设备的供电系统进行了测试,并被植入距离外部电源15至30mm的位置。它允许识别最关键的参数。在21mm处达到了34%的令人满意的效率,并倾向于验证所提出的设计程序。