BACKGROUND—Colonoscopy remains technically difficult in 10-20% of procedures due to variable colonic anatomy and fixation. The ability to vary endoscope shaft flexibility may help insertion to the caecum.
METHODS—Consecutive patients attending for day case colonoscopy were randomised to examination with either the conventional Olympus CF200HL (200HL) or a new variable stiffness (VS) colonoscope. Intubation time, use of stiffening function, and patient pain scores were compared.
RESULTS—Of 100 cases, 43 were performed with the 200HL and 57 with the VS. Four incomplete examinations occurred with the 200HL (two sigmoid fixations, two benign strictures) and two with the VS (one obstructing cancer, one fixed sigmoid). Changing to the paediatric scope was successful in all but one patient from each group (obstructive lesions). Stiff mode was applied 23 times in 18 patients and was effective in 15 of these. Intubation time was quicker with the VS (median 6 minutes 32 seconds) than with the 200HL (median 10 minutes 35 seconds) (p=0.0005). Pain scores were less with the VS (median 7) than with the 200HL (median 24) (p=0.0081).
CONCLUSIONS—The variable stiffness colonoscope combines paediatric shaft characteristics with the ability to stiffen when needed. This instrument significantly reduces intubation time and patient discomfort. Further comparisons should be made with the newest colonoscopes which are less stiff.


>Keywords: colonoscopy; colonoscopes; technology; pain
机译:背景技术由于结肠解剖结构和固定方式的变化,结肠镜检查在10-20%的手术中仍然存在技术难度。改变内窥镜杆柔韧性的能力可能有助于盲肠的插入。
方法-连续进行日间结肠镜检查的连续患者被随机分配到常规Olympus CF200HL(200HL)或新的刚度可变(VS)结肠镜检查。比较了插管时间,加强功能的使用和患者的疼痛评分。
结果-100例中,200HL进行了43例,VS进行了57例。 200HL(两次乙状结肠固定,两次良性狭窄)发生四次不完全检查,而VS(两次阻塞性癌症,乙状结肠固定一次)发生四次不完全检查。除一组外,其他所有患者(阻塞性病变)均成功改用儿科手术。僵硬模式在18例患者中使用了23次,其中15例有效。 VS(中位数6分钟32秒)的插管时间比200HL(中位数10 分钟35秒)的插管时间更快(p = 0.0005)。 VS(中位数7)的疼痛评分低于200HL(中位数24)的疼痛评分(p = 0.0081)。
结论—可变刚度结肠镜结合了儿科杆身的特点和在需要时变硬的能力。该仪器显着减少了插管时间和患者不适。应当使用刚度较小的最新结肠镜进行进一步的比较。