Breakage of the Tail Portion of the Lag Screw during Removal of Proximal Femoral Zimmer Natural Nail: Report of Two Cases with Technical Notes

机译:股骨近端自然钉去除过程中滞后螺钉尾部的断裂:附技术注释的两例报告

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Proximal femoral cephalomedullary nailing is one of the treatments of choice for intertrochanteric fracture. Zimmer Natural Nail® (ZNN; Zimmer) is one of the alternatives. We report two cases of broken tail portion of the ZNN lag screw during the removal procedure. This may be the first reported cases in scientific literature with this pattern of failure. We report these cases to highlight one of the possible problems that could occur during removal of this implant which can possibly prolong the operation time significantly if the required device is not prepared.
机译:股骨近端头髓髓钉是股骨粗隆间骨折的首选治疗方法之一。 Zimmer NaturalNail®(ZNN; Zimmer)是替代品之一。我们报告了在移除过程中ZNN尾钉的尾部断裂的两种情况。这可能是科学文献中首次报道这种失败模式的案例。我们报告这些情况,以突出在移除此植入物期间可能发生的可能问题之一,如果未准备好所需的设备,则可能会显着延长手术时间。