首页> 美国卫生研究院文献>Indian Heart Journal >Medial subclavicular musculotendinous complex and insulation break: Rare cause of late pacemaker lead malfunction
【2h】

Medial subclavicular musculotendinous complex and insulation break: Rare cause of late pacemaker lead malfunction

机译:锁骨下肌腱膜内侧复合物和绝缘破坏:起搏器晚期导线故障的罕见原因

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Insulation break in a permanent pacemaker lead is a rare long-term complication. We describe an elderly male with a VVIR pacemaker, who presented with an episode of presyncope more than 3 years after the initial implantation procedure, attributed to insulation break possibly caused by lead entrapment in components of the medial subclavicular musculotendinous complex (MSMC) and repeated compressive damage over time during ipsilateral arm movement requiring lead replacement. The differential diagnosis of a clinical presentation when pacing stimuli are present with failure to capture and the role of the MSMC in causing lead damage late after implantation are discussed.
机译:永久性起搏器导线中的绝缘断裂是一种罕见的长期并发症。我们描述了一名带有VVIR起搏器的老年男性,该男性在初始植入手术后3年以上出现晕厥前发作,这归因于绝缘破裂可能是由锁骨下锁骨下肌腱膜复合物(MSMC)的成分中夹带铅和反复压迫在同侧手臂运动过程中,随着时间的推移会造成损坏,需要更换导线。讨论了起搏刺激存在但无法捕获的临床表现的鉴别诊断,以及植入后晚期MSMC在导致铅损伤中的作用。