首页> 美国卫生研究院文献>Indian Pacing and Electrophysiology Journal >An unusual cause of recurrent syncope in a patient with implantable cardioverter defibrillator
【2h】

An unusual cause of recurrent syncope in a patient with implantable cardioverter defibrillator

机译:植入式心脏复律除颤器患者反复晕厥的异常原因

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Electromagnetic interferences (EMI) deriving from electrical devices may affect implantable cardioverter defibrillators (ICD). Improved algorithms have been developed in order to minimize adverse effects. However, caution should be still recommended in ICD recipients when handling electrical devices. Here we describe the case of an ICD patient with recurrent syncopal episodes due to inhibition of pacing by oversensing of electrical noise from a not properly grounded washing machine.
机译:来自电子设备的电磁干扰(EMI)可能会影响植入式心脏复律除颤器(ICD)。为了最小化不利影响,已经开发了改进的算法。但是,在处理电子设备时,仍应建议ICD接收者注意。在这里,我们描述了由于未正确接地的洗衣机对电噪声的过度感应而抑制起搏,导致反复发作晕厥发作的ICD患者。