Multiplex Detection of IgM and IgG Class Antibodies to Toxoplasma gondii Rubella Virus and Cytomegalovirus Using a Novel Multiplex Flow Immunoassay

机译:使用新型多重流免疫测定法对弓形虫风疹病毒和巨细胞病毒的IgM和IgG类抗体进行多重检测

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The goal of this study was to evaluate the BioPlex 2200 Toxoplasma, rubella, and cytomegalovirus (CMV) (ToRC) IgG and IgM multiplex immunoassays (Bio-Rad Laboratories, Hercules, CA) and compare the results to those of conventional testing by enzyme immunoassay (EIA) and enzyme-linked fluorescent assay (ELFA). Serum specimens (n = 600) submitted for routine ToRC IgG and IgM testing by EIA (SeraQuest, Doral, FL; Diamedix, Miami, FL) or ELFA (Vidas; bioMérieux, Durham, NC) were also tested by the BioPlex ToRC multiplex immunoassays. Samples showing discordant results were retested by both methods, with further discrepancies being arbitrated by a third assay. Following repeat testing, the BioPlex Toxoplasma, rubella, and CMV IgG assays demonstrated agreements of 98.7 (592/600 specimens), 93.3 (560/600 specimens), and 98.3% (590/600 specimens), respectively, while the ToRC IgM assays yielded agreements of 91.2 (547/600 specimens), 87.3 (524/600 specimens), and 95.2% (571/600 specimens), respectively. The BioPlex ToRC IgG assays provided results comparable to EIA/ELFA results, with kappa coefficients showing near-perfect agreement for the Toxoplasma (κ = 0.94) and CMV (κ = 0.97) IgG assays and substantial agreement for the rubella IgG assay (κ = 0.66). The BioPlex ToRC IgM assays showed lower specificity with only slight agreement for Toxoplasma IgM (κ = 0.07), poor agreement for rubella IgM (κ = −0.03), and moderate agreement for CMV IgM (κ = 0.55). Both the BioPlex IgG and IgM assays reduced turnaround time (1.7 h versus 5.5 h by EIA/ELFA for 100 specimens) and eliminated the necessity to manually pipette or aliquot specimens prior to testing.
机译:这项研究的目的是评估BioPlex 2200弓形虫,风疹和巨细胞病毒(CMV)(ToRC)IgG和IgM多重免疫测定(Bio-Rad Laboratories,Hercules,CA),并将结果与​​酶免疫测定的常规检测结果进行比较(EIA)和酶联荧光测定(ELFA)。还通过BioPlex ToRC多重免疫测定法对通过EIA(SeraQuest,Doral,FL; Diamedix,Miami,FL)或ELFA(Vidas;bioMérieux,Durham,NC)提交用于常规ToRC IgG和IgM检测的血清标本(n = 600)进行了测试。 。两种方法均对显示不一致结果的样品进行了重新测试,而进一步的差异则通过第三种方法进行仲裁。经过重复测试,BioPlex弓形虫,风疹和CMV IgG测定分别显示出98.7(592/600标本),93.3(560/600标本)和98.3%(590/600标本)的一致性,而ToRC IgM测定分别产生91.2(547/600标本),87.3(524/600标本)和95.2%(571/600标本)的一致性。 BioPlex ToRC IgG分析提供的结果可与EIA / ELFA结果相媲美,kappa系数显示弓形虫(κ= 0.94)和CMV(κ= 0.97)IgG分析接近完美一致性,风疹IgG分析(κ= 0.66)。 BioPlex ToRC IgM分析显示较低的特异性,仅与弓形虫IgM一致(κ= 0.07),对风疹IgM一致性差(κ= -0.03),对CMV IgM一致性中等(κ= 0.55)。 BioPlex IgG和IgM分析均减少了周转时间(对于100个样本,EIA / ELFA减少了1.7小时,而EIA / ELFA为5.5小时),并且消除了在测试之前手动吸取或分装样本的必要性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号