首页> 美国卫生研究院文献>British Medical Journal >Rapid measurement of creatine kinase activity in a coronary care unit using a portable benchtop reflectance photometer.
【2h】

Rapid measurement of creatine kinase activity in a coronary care unit using a portable benchtop reflectance photometer.

机译:使用便携式台式反射光度计快速测量冠状动脉护理单元中的肌酸激酶活性。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Rapid measurements of plasma creatine kinase activity using an inexpensive benchtop reflectance photometer (Ames Seralyzer) and disposable reagent strips were evaluated in the laboratory and coronary care unit. The system proved simple to use and capable of yielding rapid (four minutes per analysis), precise (coefficient of variation less than 9%), and accurate results (correlation with routine method 0.995) when used by medical staff. Creatine kinase values were available 6.5-102 hours earlier than routine laboratory data, depending on the time and day of sampling, thereby facilitating appropriate and economic patient management. This instrument might be used to supplement the routine enzyme service for selected admissions, resulting in greatly improved availability of results and hence contributing to the early discharge of patients from intensive care facilities.
机译:使用廉价的台式反射光度计(Ames Seralyzer)和一次性试剂带对血浆肌酸激酶活性进行快速测量,并在实验室和冠心病监护室进行了评估。该系统使用简单,在医务人员使用时能够快速(每次分析四分钟),精确(变异系数小于9%)和准确结果(与常规方法0.995相关)。肌酸激酶值可比常规实验室数据早6.5-102小时获得,这取决于采样的时间和日期,从而有利于适当而经济的患者管理。该仪器可用于补充某些入院患者的常规酶服务,从而大大提高结果的可利用性,从而有助于患者从重症监护设施中早日出院。