During the initial testing of Radio Leicester a swept-frequency technique for testing radio antennae was shown to affect demand pacemakers by inhibition of the pacing impulse and to interfere with physiological monitoring equipment. Adequate filtering of demand pacemakers is necessary to eliminate this interference. There is no evidence that such testing has any effect on the function of fixed-rate pacemakers. There is also a potential danger to implanted demand pacemakers by the use of similar polyscope generators used for servicing radio and television apparatus in industry.
机译:在莱斯特广播电台的初始测试中,一种用于测试无线电天线的扫频技术被证明通过抑制起搏脉冲而影响需求的起搏器并干扰生理监测设备。需要对需求起搏器进行充分过滤,以消除这种干扰。没有证据表明这种测试对固定价格的起搏器的功能有任何影响。使用类似的多视镜发生器为工业无线电和电视设备提供服务,对于植入需求的起搏器也存在潜在的危险。