Portable spectral Doppler echocardiographic device: overcoming limitations

机译:便携式频谱多普勒超声心动图仪:克服局限性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Background: There is evidence that new portable echocardiographic devices are useful in evaluating heart anatomy and function, but a lack of Doppler modes has up to now been an important limitation in obtaining haemodynamic data.>Objectives: To report the Doppler capabilities of a new hand held echocardiographic device.>Design: Blinded comparison of two types of echocardiography machine.>Setting: Tertiary care centre.>Patients: 98 consecutive patients were randomly imaged with the hand held device, with a standard platform as reference.>Outcome measures: Pulsed wave transmitral Doppler inflow tract velocities, deceleration time, and continuous wave Doppler measurements of aortic ejection and tricuspid regurgitation peak velocities were recorded.>Results: There was excellent agreement between the hand held device and standard echocardiography for the evaluation of diastolic E and A waves, E/A ratio, and deceleration time with pulsed wave Doppler (intraclass correlation coefficients of 0.97, 0.93, 0.90, and 0.78, respectively). In addition, good agreement was found between continuous wave Doppler measurements of aortic ejection and tricuspid regurgitation velocities (intraclass correlation coefficients of 0.96 and 0.80). However, there was a significant difference between patients with tricuspid regurgitation measured with the hand held device (25.5%) and by standard echocardiography (65.3%), resulting in misdiagnosis of eight patients with pronounced pulmonary hypertension.>Conclusions: New hand held devices with Doppler capabilities overcome previous limitations in evaluating haemodynamic variables. With colour Doppler they are now suitable for the complete evaluation of valvar disease and diastolic function. However, important limitations remain in the evaluation of pulmonary pressures.
机译:>背景:有证据表明,新型便携式超声心动图设备可用于评估心脏的解剖结构和功能,但是到目前为止,缺乏多普勒模式一直是获取血液动力学数据的重要限制。>目的: 报告新型手持超声心动图设备的多普勒功能。>设计:两种超声心动图设备的盲目比较。>设置:三级医疗中心。 >患者:使用标准平台作为参考,使用手持设备对连续98位患者进行随机成像。>结果测量:脉冲波透射多普勒血流速度,减速时间和连续波记录多普勒测量的主动脉射血和三尖瓣关闭不全峰值速度。>结果:手持式设备与标准超声心动图在评估舒张期E方面有很好的一致性A波,E / A比和带脉冲多普勒的减速时间(类内相关系数分别为0.97、0.93、0.90和0.78)。此外,在连续波多普勒测量主动脉射血与三尖瓣关闭不全速度之间也发现了很好的一致性(类内相关系数为0.96和0.80)。但是,使用手持式设备(25.5%)和标准超声心动图(65.3%)测量的三尖瓣关闭不全患者之间存在显着差异,导致误诊了8例明显的肺动脉高压患者。>结论:具有多普勒功能的新型手持设备克服了先前在评估血流动力学变量方面的局限性。借助彩色多普勒仪,它们现在适合全面评估瓣膜疾病和舒张功能。但是,在评估肺压方面仍然存在重要的局限性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号