Estimation of cardiac reserve by peak power: validation and initial application of a simplified index

机译:通过峰值功率估算心脏储备:简化指标的验证和初步应用

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

OBJECTIVES—To validate a simplified estimate of peak power (SPP) against true (invasively measured) peak instantaneous power (TPP), to assess the feasibility of measuring SPP during exercise and to correlate this with functional capacity.
DESIGN—Development of a simplified method of measurement and observational study.
SETTING—Tertiary referral centre for cardiothoracic disease.
SUBJECTS—For validation of SPP with TPP, seven normal dogs and four dogs with dilated cardiomyopathy were studied. To assess feasibility and clinical significance in humans, 40 subjects were studied (26 patients; 14 normal controls).
METHODS—In the animal validation study, TPP was derived from ascending aortic pressure and flow probe, and from Doppler measurements of flow. SPP, calculated using the different flow measures, was compared with peak instantaneous power under different loading conditions. For the assessment in humans, SPP was measured at rest and during maximum exercise. Peak aortic flow was measured with transthoracic continuous wave Doppler, and systolic and diastolic blood pressures were derived from brachial sphygmomanometry. The difference between exercise and rest simplified peak power (Δ SPP) was compared with maximum oxygen uptake (V̇O2max), measured from expired gas analysis.
RESULTS—SPP estimates using peak flow measures correlated well with true peak instantaneous power (r = 0.89 to 0.97), despite marked changes in systemic pressure and flow induced by manipulation of loading conditions. In the human study, V̇O2max correlated with Δ SPP (r = 0.78) better than Δ ejection fraction (r = 0.18) and Δ rate-pressure product (r = 0.59).
CONCLUSIONS—The simple product of mean arterial pressure and peak aortic flow (simplified peak power, SPP) correlates with peak instantaneous power over a range of loading conditions in dogs. In humans, it can be estimated during exercise echocardiography, and correlates with maximum oxygen uptake better than ejection fraction or rate-pressure product.


Keywords: stress echocardiography; oxygen consumption; left ventricular function; cardiac power output
机译:目标—相对于真实(有创测量)峰值瞬时功率(TPP)验证峰值功率(SPP)的简化估计,评估在运动过程中测量SPP的可行性并将其与功能能力相关联。
设计—开发测量和观察方法的简化方法。
设置-心胸疾病三级转诊中心。
主题-为了验证TPP对SPP的有效性,研究了七只正常犬和四只扩张型心肌病犬。为了评估在人类中的可行性和临床意义,研究了40位受试者(26位患者; 14位正常对照)。
方法-在动物验证研究中,TPP来源于升主动脉压和流量探针以及多普勒测量流。将使用不同流量测量方法计算出的SPP与不同负载条件下的峰值瞬时功率进行比较。为了对人体进行评估,在休息和最大运动量时对SPP进行了测量。用经胸连续波多普勒测量主动脉血流量峰值,并从肱血压计得出收缩压和舒张压。将运动和休息的简化峰值功率(Δ SPP)之间的差异与最大气体摄入量(V̇O2max)进行了比较,这是根据过期气体分析测得的。具有真正的峰值瞬时功率(r = 0.89至0.97),尽管通过操纵负载条件引起的系统压力和流量有明显变化。在人体研究中,V̇O2max与ΔSPP(r = 0.78)的相关性优于Δ射血分数(r = 0.18)和Δ率压乘积(r = 0.59)。
结论-平均动脉压的简单乘积在犬的一系列负荷条件下,主动脉血流峰值(简化的峰值功率,SPP)与峰值瞬时功率相关。在人类中,可以在运动超声心动图期间对其进行估计,并且与最大摄氧量的相关性要好于射血分数或速率-压力乘积。


耗氧量左心室功能;心力输出