Effect of Low Refocusing Angle in T1-Weighted Spin Echo and Fast Spin Echo MRI on Low-Contrast Detectability: A Comparative Phantom Study at 1.5 and 3 Tesla

机译:T1加权自旋回波和快速自旋回波MRI中低重聚焦角度对低对比度可检测性的影响:在1.5和3特斯拉的比较幻像研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

MRI tissue contrast is not well preserved at high field. In this work, we used a phantom with known, intrinsic contrast (3.6%) for model tissue pairs to test the effects of low angle refocusing pulses and magnetization transfer from adjacent slices on intrinsic contrast at 1.5 and 3 Tesla. Only T1-weighted spin echo sequences were tested since for such sequences the contrast loss, tissue heating, and image quality degradation at high fields seem to present significant diagnostic and quality issues. We hypothesized that the sources of contrast loss could be attributed to low refocusing angles that do not fulfill the Hahn spin echo conditions or to magnetization transfer effects from adjacent slices in multislice imaging. At 1.5 T the measured contrast was 3.6% for 180° refocusing pulses and 2% for 120° pulses, while at 3 T, it was 4% for 180° and only 1% for 120° refocusing pulses. There was no significant difference between single slice and multislice imaging suggesting little or no role played by magnetization transfer in the phantom chosen. Hence, one may conclude that low angle refocusing pulses not fulfilling the Hahn spin echo conditions are primarily responsible for significant deterioration of T1-weighted spin echo image contrast in high-field MRI.
机译:MRI组织的对比度在高视野下不能很好地保留。在这项工作中,我们对模型组织对使用了具有已知固有对比度(3.6%)的模型,以测试低角度重新聚焦脉冲和相邻切片的磁化转移对1.5和3 Tesla固有对比度的影响。仅测试了T1加权的自旋回波序列,因为对于此类序列,高场下的对比度损失,组织发热和图像质量下降似乎带来了重大的诊断和质量问题。我们假设对比度损失的原因可能归因于不能满足Hahn自旋回波条件的低重聚焦角度,或者归因于多层成像中相邻切片的磁化转移效应。在1.5 T下,180°重聚焦脉冲的测量对比度为3.6%,在120°脉冲下为2%,而在3 T下,180°重聚焦脉冲为4%,而120°重聚焦脉冲仅为1%。单层和多层成像之间没有显着差异,表明在所选模型中磁化转移几乎没有作用。因此,可以得出结论,不满足哈恩自旋回波条件的低角度重聚焦脉冲是造成高场MRI中T1加权自旋回波图像对比度显着下降的主要原因。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号