Temperature Values Variability in Piezoelectric Implant Site Preparation: Differences between Cortical and Corticocancellous Bovine Bone

机译:压电植入位点准备中的温度值变异性:皮质和皮质tic骨牛骨之间的差异

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Purpose. Various parameters can influence temperature rise and detection during implant site preparation. The aim of this study is to investigate local temperature values in cortical and corticocancellous bovine bone during early stages of piezoelectric implant site preparation. Materials and Methods. 20 osteotomies were performed using a diamond tip (IM1s, Mectron Medical Technology, Carasco, Italy) on two different types of bovine bone samples, cortical and corticocancellous, respectively. A standardized protocol was designed to provide constant working conditions. Temperatures were measured in real time at a fixed position by a fiber optic thermometer. Results. Significantly higher drilling time (154.90 sec versus 99.00 sec; p < 0.0001) and temperatures (39.26°C versus 34.73°C; p = 0.043) were observed in the cortical group compared to the corticocancellous group. A remarkable variability of results characterized the corticocancellous blocks as compared to the blocks of pure cortical bone. Conclusion. Bone samples can influence heat generation during in vitro implant site preparation. When compared to cortical bone, corticocancellous samples present more variability in temperature values. Even controlling most experimental factors, the impact of bone samples still remains one of the main causes of temperature variability.
机译:目的。各种参数会影响植入部位准备过程中的温度上升和检测。这项研究的目的是调查在压电植入部位准备的早期阶段皮质和皮质小牛的骨中的局部温度值。材料和方法。使用金刚石针头(IM1s,Mectron Medical Technology,Carasco,意大利)对两种不同类型的牛骨样品分别进行了20处截骨术,分别是皮层皮层和皮层皮层。设计了标准化协议以提供恒定的工作条件。通过光纤温度计在固定位置实时测量温度。结果。与皮质囊肿组相比,皮质组观察到明显更高的钻孔时间(154.90 sec对99.00 sec; p <0.0001)和温度(39.26℃对34.73℃; p = 0.043)。与纯皮质骨的块相比,结果显着的变化是皮质小结块的特征。结论。骨样本可能会影响体外植入部位准备过程中的热量产生。当与皮质骨相比时,皮质囊样样品在温度值上表现出更大的可变性。即使控制大多数实验因素,骨骼样品的影响仍然仍然是温度变化的主要原因之一。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号