首页> 美国卫生研究院文献>BioMed Research International >Monitoring the Changes of Material Properties at Bone-Implant Interface during the Healing Process In Vivo: A Viscoelastic Investigation
【2h】

Monitoring the Changes of Material Properties at Bone-Implant Interface during the Healing Process In Vivo: A Viscoelastic Investigation

机译:体内修复过程中监测骨-植入物界面材料性质的变化:粘弹性研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

The aim of this study was to monitor the changes of viscoelastic properties at bone-implant interface via resonance frequency analysis (RFA) and the Periotest device during the healing process in an experimental rabbit model. Twenty-four dental implants were inserted into the femoral condyles of rabbits. The animals were sacrificed immediately after implant installation or on day 14, 28, or 56 after surgery. Viscoelastic properties at bone-implant interface were evaluated by measuring the implant stability quotient (ISQ) using RFA and by measuring the Periotest values (PTVs) using the Periotest device. The bone/implant specimens were evaluated histopathologically and histomorphometrically to determine the degree of osseointegration (BIC%). The BIC% values at different time points were then compared with the corresponding ISQ values and PTVs. The mean ISQ value increased gradually and reached 81 ± 1.7 on day 56, whereas the mean PTV decreased over time, finally reaching −0.7 ± 0.5 on day 56. Significant correlations were found between ISQ and BIC% (r = 0.701, p < 0.001), PTV and BIC% (r = −0.637, p < 0.05), and ISQ and PTV (r = −0.68, p < 0.05). These results show that there is a positive correlation between implant stability parameters and peri-implant-bone healing, indicating that the RFA and Periotest are useful for measuring changes of viscoelastic properties at bone-implant interface and are reliable for indirectly predicting the degree of osseointegration.
机译:这项研究的目的是在实验兔模型的愈合过程中,通过共振频率分析(RFA)和Periotest装置监测骨-植入物界面的粘弹性特性的变化。将二十四颗牙齿植入物插入兔子的股骨dy中。在植入物安装后或手术后第14、28或56天立即处死动物。通过使用RFA测量植入物稳定性商(ISQ)和使用Periotest装置测量Periotest值(PTV),可以评估骨-植入物界面的粘弹性。对骨/植入物标本进行组织病理学和组织形态学评估,以确定骨整合程度(BIC%)。然后将不同时间点的BIC%值与相应的ISQ值和PTV进行比较。平均ISQ值逐渐增加,在第56天达到81±1.7,而平均PTV随着时间的流逝而下降,最终在第56天达到-0.7±0.5。ISQ与BIC%之间存在显着相关性(r = 0.701,p <0.001) ),PTV和BIC%(r = -0.637,p <0.05),以及ISQ和PTV(r = -0.68,p <0.05)。这些结果表明,植入物稳定性参数与植入物周围骨愈合之间存在正相关,表明RFA和Periotest可用于测量骨-植入物界面处的粘弹性特性的变化,并且可以间接地预测骨整合的程度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号