首页> 美国卫生研究院文献>Biochemia Medica >Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices
【2h】

Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices

机译:庆大霉素检测中未发现明显的正负变化:对常规实验室质量操作的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

IntroductionA product recall was issued for the Roche/Hitachi Cobas Gentamicin II assays on 25th May 2016 in Australia, after a 15 - 20% positive analytical shift was discovered. Laboratories were advised to employ the Thermo Fisher Gentamicin assay as an alternative. Following the reintroduction of the revised assay on 12th September 2016, a second reagent recall was made on 20th March 2017 after the discovery of a 20% negative analytical shift due to erroneous instrument adjustment factor.
机译:简介在发现15-20%的阳性分析变化后,于2016年5月25日在澳大利亚发布了针对Roche / Hitachi Cobas Gentamicin II测定的产品召回。建议实验室采用赛默飞世尔庆大霉素测定法作为替代方法。在2016年9月12日重新引入修订的检测方法后,由于发现了20%的负分析偏移,在2017年3月20日再次召回了试剂仪器调整系数错误。