We report on a preanalytical issue we encountered during routine clinical chemistry analyses, potentially leading to deviated analysis results and believe that it might help other laboratories to overcome similar problems. In a heparin-gel tube we measured an implausible glucose value of 0.06 mmol/L. Re-measurement of the same sample resulted in a glucose value of 5.4 mmol/L. After excluding an analytical error, we inspected the sample closer and found a white material as well as fatty droplets floating on the surface of the plasma tube. Evaluation of these structures revealed that the white particulate matter (WPM) consisted of fibrinogen, platelets and leukocytes and the fatty droplets most probably originated from the separator gel. We concluded that these structures formed a temporary clot in the instruments pipetting needle thereby altering the sampling volume and subsequently the measured glucose value. The formation of WPM might be attributable to high speed centrifugation, high cholesterol levels, the gel formulation or a combination of several issues such as temperature, heparin concentration, pH and patient-specific factors. The gel droplets were most probably caused by an aberrant gel formulation in combination with an improper storage of the empty tubes on the wards prior to phlebotomy. After adding an additional instrument cleansing cycle and changing to another batch of heparin tubes the problems could be significantly reduced.
机译:我们报告了在常规临床化学分析过程中遇到的分析前问题,有可能导致分析结果出现偏差,并相信这可能有助于其他实验室克服类似问题。在肝素凝胶管中,我们测得的葡萄糖值为0.06 mmol / L。再次测量同一样品的葡萄糖值为5.4 mmol / L。排除分析误差后,我们仔细检查了样品,发现白色物质以及等离子管表面上漂浮的脂肪滴。对这些结构的评估表明,白色颗粒物(WPM)由纤维蛋白原,血小板和白细胞组成,而脂肪滴最有可能来自分离器凝胶。我们得出的结论是,这些结构在仪器的移液针中形成了一个临时凝块,从而改变了采样量,进而改变了测得的葡萄糖值。 WPM的形成可能归因于高速离心,高胆固醇水平,凝胶制剂或多种问题的组合,例如温度,肝素浓度,pH和患者特异性因素。凝胶滴很可能是由异常的凝胶配方与静脉放血前病房中空管的不适当存储引起的。添加了额外的仪器清洁周期并更换为另一批肝素管后,可以大大减少问题。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号