Important objectives of External Quality Assessment (EQA) is to detect analytical errors and make corrective actions. The aim of this paper is to describe knowledge required to interpret EQA results and present a structured approach on how to handle deviating EQA results. The value of EQA and how the EQA result should be interpreted depends on five key points: the control material, the target value, the number of replicates, the acceptance limits and between lot variations in reagents used in measurement procedures. This will also affect the process of finding the sources of errors when they appear. The ideal EQA sample has two important properties: it behaves as a native patient sample in all methods (is commutable) and has a target value established with a reference method. If either of these two criteria is not entirely fulfilled, results not related to the performance of the laboratory may arise. To help and guide the laboratories in handling a deviating EQA result, the Norwegian Clinical Chemistry EQA Program (NKK) has developed a flowchart with additional comments that could be used by the laboratories e.g. in their quality system, to document action against deviations in EQA. This EQA-based trouble-shooting tool has been developed further in cooperation with the External quality Control for Assays and Tests (ECAT) Foundation. This flowchart will become available in a public domain, i.e. the website of the European organisation for External Quality Assurance Providers in Laboratory Medicine (EQALM).
机译:外部质量评估(EQA)的重要目标是发现分析错误并采取纠正措施。本文的目的是描述解释EQA结果所需的知识,并提出有关如何处理偏离EQA结果的结构化方法。 EQA的值以及如何解释EQA的结果取决于五个关键点:对照材料,目标值,重复次数,接受极限以及在测量程序中使用的试剂批次之间的变化。这也将影响错误源的发现过程。理想的EQA样本具有两个重要属性:在所有方法中均表现为患者样本(可互换),并具有通过参考方法确定的目标值。如果这两个标准中的任何一个都不完全满足,则可能会产生与实验室性能无关的结果。为了帮助和指导实验室处理偏离EQA的结果,挪威临床化学EQA计划(NKK)制定了流程图,并附有可供实验室使用的其他注释,例如在其质量体系中,记录针对EQA偏差采取的措施。该基于EQA的故障排除工具已与“分析和测试外部质量控制”基金会(ECAT)基金会进一步合作开发。该流程图将在公共领域,即欧洲实验室医学外部质量保证提供者组织(EQALM)的网站上提供。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号