Registry study on failure incidence in 1127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve

机译:随访10年后对1127颗经改良的髋关节植入物并再次使用茎耳轴的失败发生率进行了登记研究:适配器套的影响有限

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background and purpose — Little is known about the role of retained trunnions in revision hip arthroplasties, i.e., when only the femoral head is substituted. Wear (fretting corrosion) and ceramic head fractures are 2 poorly understood concerns related to use, and the role of adapter sleeves has not been defined. In this registry study we assessed the influence of sleeve interposition on re-revision rates in revision hip arthroplasties with retained stems. Confounding factors (demographics, implant-related features) and failures were also analyzed.Patients and methods — We conducted a registry study on 1,127 revised implants (retained trunnion and head exchange). In 26% of implants an adapter sleeve was interposed; in 74% no adapter sleeve was implanted. Demographic and implant-related features were investigated including a descriptive analysis of failures.Results — The mean follow-up of revised implants with and without the use of an adapter sleeve was 3.3 and 5.1 years, respectively. The implant survival without an adapter sleeve was significantly higher, 98.4% (95% CI 96.9–99.8) vs. 95.2% (CI 93.2–96.6) with an adapter sleeve at 5 years. No re-revisions due to adverse local tissue reactions or ceramic head fractures were reported. In order to overcome the different distribution of head materials and head sizes in the two cohorts, only Delta balls were investigated.Interpretation — Adapter sleeve interposition had a minor influence on the revision rates. No adverse local tissue reactions or head fractures occurred.
机译:背景和目的-对于保留的耳轴在翻修髋关节置换术中(即仅用股骨头置换)的作用了解甚少。磨损(微动腐蚀)和陶瓷头断裂是与使用有关的2个鲜为人知的关注点,转接套筒的作用尚未定义。在这项注册表研究中,我们评估了在保留茎的修订型髋关节置换术中,套管插入对再修订率的影响。还对混杂因素(人口统计学,与植入物相关的特征)和失败进行了分析。患者和方法—我们对1,127枚改良的植入物(保留的耳轴和头部置换)进行了注册表研究。在26%的植入物中插入了紧定套。 74%的患者未植入任何适配套筒。研究了人口统计学特征和与植入物有关的特征,包括对失败的描述性分析。结果—在使用和不使用紧定套的情况下,改良植入物的平均随访时间分别为3.3年和5.1年。不使用紧定套的植入物在5年时的存活率要高得多,达到98.4%(95%CI 96.9–99.8),而使用紧定套的植入物生存率为95.2%(CI 93.2–96.6)。没有报告由于不良的局部组织反应或陶瓷头骨折而进行的修订。为了克服两个队列中头部材料和头部大小的不同分布,仅研究了Delta球。解释—紧定套的插入对改版率影响较小。没有发生不利的局部组织反应或头部骨折。

著录项

相似文献

  • 外文文献
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号