摘要

AIM:To evaluate the clinical efficacy of an expanded polytetrafluoro-ethylene-covered Fluency stent compared with that of a polyethylene terephthalate-covered Wallgraft stent for the management of transjugular intrahepatic portosystemic shunt(TIPS)dysfunction.METHODS:A retrospective review of patients who underwent TIPS revision with stent-grafts between May 2007 and June 2011 was conducted.The patients were divided into two groups according to the stentgrafts implanted:the Fluency stent(Bard Incorporated,Karlsruhe,Germany)and the Wallgraft stent(Boston Scientific,Galway,Ireland).The primary patency rates were calculated and compared using the Kaplan-Meier method.RESULTS:A total of 73 patients were evaluated in this study:33 with Fluency stents and 40 with Wallgraft stents.The primary patency rates at 12 and 24 mo were 91% and 85%,respectively,in the Fluency stent group and 78% and 63%,respectively,in the Wallgraft stent group.The primary shunt patency rates after TIPS revision were significantly better with the Fluency stent than with the Wallgraft stent(P = 0.033).CONCLUSION:TIPS revision with the Fluency stent has higher medium-term patency rates than that with the Wallgraft stent.