首页> 中文期刊> 《中德临床肿瘤学杂志(英文版)》 >A dosimetric evaluation of flattening filter-free volumetric modulated arc therapy for postoperative treatment of cervical cancer

A dosimetric evaluation of flattening filter-free volumetric modulated arc therapy for postoperative treatment of cervical cancer

         

摘要

Objective The aim of the study was to compare flattening filter-free(FFF) beams and conventional flattening filter(FF) beams in volumetric modulated arc therapy(VMAT) for cervical cancer after surgery, through a retrospective planning study.Methods VMAT plans of FFF beams and normal FF beams were designed for a cohort of 15 patients. The prescribed dose was 45 Gy to 1.8 Gy per fraction, and at least 95% of the planning target volume received this dose. Doses were computed with a commercially available treatment planning system using a Monte Carlo(MC) algorithm. Plans were compared according to dose-volume histogram analysis in terms of planning target volume homogeneity and conformity indices(HI and CI), as well as organs at risk(OAR) dose and volume parameters. Results FFF-VMAT was similar to FF-VMAT in terms of CI, but inferior to FF-VMAT considering HI. No statistically differences were observed between FFF-VMAT and FF-VMAT in following organ at risks including pelvic bone marrow, small bowel, bladder, rectum, and normal tissue(NT)..Conclusion For patients with cervical cancer after hysterectomy, the FFF beam achieved target and OAR dose distribution similar to that of the FF beam. Reduction of beam-on time in cervical cancer is beneficial.