We propose an in-situ method to calibrate the coil constants of the optical atomic magnetomete. This method is based on measuring the Larmor precession of spin polarized alkali metal atoms and has been demonstrated on a K-Rb hybrid atomic magnetomete. Oscillation fields of different frequencies are swept on the transverse coil. By extracting the resonance frequency through phase-frequency analysis of electron spin projection, the coil constants are calibrated to be 323.1±0.28 nT/mA, 108±0.04nT/Ma, and 185.8±1.03 nT/mA along the X,Y, and Z directions, respectively.
机译:我们提出了一种原位方法来校准光学原子磁铁的线圈常数。该方法基于测量自旋极化碱金属原子的拉莫尔进动,并已在K-Rb混合原子磁法上得到证明。在横向线圈上扫描了不同频率的振荡场。通过电子自旋投影的相频分析提取共振频率,沿X,Y和X方向将线圈常数校准为323.1±0.28 nT / mA,108±0.04nT / Ma和185.8±1.03 nT / mA。分别为Z方向。