首页> 中文期刊> 《药物分析学报:英文版》 >EFFECTS OF FILTER MEMBRANES ADSORBING INFLAMMATORY MEDIATORS IN HEMOFILTRATI ON

EFFECTS OF FILTER MEMBRANES ADSORBING INFLAMMATORY MEDIATORS IN HEMOFILTRATI ON

         

摘要

Objective In this study, an in vitro hemofiltration model was set up to investigate adsorptive saturati on time of different membrane under different blood flow rate (Q B)and filt rate rate(Q F). Methods Anticoagulated cattle blood (2000 mL per bag) was stimulated with 1μg·mL -1 endotoxin to induce i nflammatory mediators before hemofiltration (HF) using AN69, PS and PMMA filters in vitro. Adsorptive saturation time of membrane was observed using data br idge in different Q B and different Q F. TNF was measured by radioimmu noassays. Results Before the resistance level reached the peak value in the same group, resistance level increased significantly (P05). It was suggested tha t the resistance level reached plateau at 150,120,90,120, and150 minutes in Q F of 100 mL·minute -1, 200 mL·minute -1, 300 mL·minute -1 , respectively and in Q F of 1L·hour -1, 2L·hour -1, 4L·hour -1, respectively. And with the Q B and Q F increasing, resistanc e level increased significantly (P<0.01) among different groups at the same time point in A, B, C, D and E group. Conclusion Membrane resis tance level online measured by Data Bridge can instantly reflect the degree of m embrane adsorption. Adsorptive saturation time of filter membrane in different f iltration flow rate and blood flow rate are different.