首页> 外文期刊>国际眼科杂志:英文版 >Comparison of corneal endothelial changes following phacoemulsification with transversal and torsional phacoemulsification machines
【24h】

Comparison of corneal endothelial changes following phacoemulsification with transversal and torsional phacoemulsification machines

机译:横向和扭转超声乳化仪超声乳化术后角膜内皮变化的比较

获取原文
获取原文并翻译 | 示例
           

摘要

AIM:To compare and evaluate the phacoemulsification parameters and postoperative endothelial cell changes of two different phacoemulsification machines, each with different modes, but also to assess the relationship between postoperative endothelial cell loss and the phacoemulsification parameters, as well as the other factors in both groups.METHODS:This prospective observational study was comprised of consecutive eligible cataract patients operated with phacoemulsification technique performed by the same surgeon using either a WHITESTAR Signature Ellips FX(transversal, group 1) or Infiniti OZil IP(torsional, group 2) machine.RESULTS:The study included 86 patients. Baseline characteristics in the groups were similar. The median nuclear sclerosis grade was 3(2-4) in the first group and2(2-4) in the second group(P =0.265). Both groups had similar phacoemulsification needle times(group 1: 60.63±36 s; group 2: 55.98±30 s; P =0.789). The percentage of endothelial cell loss 30 d after surgery ranged from 3% to15% with a median of 7% in group 1, and from 2% to13% with a median of 6% in group 2; however, there was no statistically significant difference between the groups(P =0.407). Hexagonality(P =0.794) and the coefficient of variation(CV; P =0.142) did not differ significantly between the groups before and 30 d after surgery. Asignificant positive correlation was found between the endothelial cell loss and nuclear sclerosis grade(group1: P 0.001; group 2: P 0.001) and between the endothelial cell loss and average phacoemulsification power(group 1: P =0.007; group 2: P =0.008).CONCLUSION:Both of these machines were efficient,with similar endothelial cell loss. This endothelial cell loss was related to the increased nuclear sclerosis grade and increased phacoemulsification power.
机译:目的:比较和评估两种不同超声乳化机的超声乳化参数和术后内皮细胞的变化,每种超声乳化机的模式各不相同,还评估术后内皮细胞损失与超声乳化参数以及两组其他因素之间的关系。方法:这项前瞻性观察研究由连续的合格白内障患者组成,这些患者由同一名外科医生使用WHITESTAR Signature Ellips FX(横向,第1组)或Infiniti OZil IP(扭转,第2组)机器进行超声乳化术进行手术。该研究包括86名患者。各组的基线特征相似。第一组的中位核硬化等级为3(2-4),第二组为2(2-4)(P = 0.265)。两组超声乳化针的时间相似(组1:60.63±36 s;组2:55.98±30 s; P = 0.789)。术后30 d,内皮细胞丢失的百分比在第1组中为3%至15%,中位数为7%;在第2组中为2%至13%,中位数为6%。然而,两组之间没有统计学差异(P = 0.407)。术前和术后30 d两组之间的六角形(P = 0.794)和变异系数(CV; P = 0.142)无显着差异。内皮细胞损失与核硬化等级之间存在显着正相关(组1:P <0.001;组2:P <0.001);内皮细胞损失与平均超声乳化力之间(组1:P = 0.007;组2:P = 0.008)。结论:这两种机器均有效,内皮细胞损失相似。内皮细胞的损失与核硬化等级的增加和超声乳化能力的增加有关。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号