采用全自动石墨消解仪对样品进行前处理,用冷原子吸收分光光度法测定样品中汞的含量.该方法在0~20 ng范围内线性良好,线性方程为y(A)=0.021+1.517x(C),相关系数r2=0.9999,检出限为0.0008 mg/kg.对实际土壤样品进行6次平行测定,相对标准偏差为0.4 %~1.1 %.对标准土壤样品进行6次平行测定,测定结果均在标准值允许范围内.%The sample was pretreated with an automatic graphite digestion instrument, and the content of mercury in the sample was determined by cold atomic absorption spectrophotometry. The method is well linear in the range of 0~20 ng, the linear equation is y (A)=0.021+1.517x(C), the correlation coefficient is r2=0.9999, and the detection limit is 0.0008 mg/kg. 6 parallel measurements were carried out on the actual soil samples, with a relative standard deviation of 0.4 %~1.1 %. 6 parallel measurements of standard soil samples were carried out, and the results were within the allowable range of standard values.