This paper reports on the development of a novel electrochemical assay for cadmium (II) in natural water, which involves the use of disposable hydroxyapatite modified platinum electrode (HAP/Pt). Cadmium (II) was preconcentrated on the surface of the modified electrode and adsorbed onto HAP and oxidized at E = –680 mV. The HAP-modified platinum electrode exhibited superior performance in comparison to the plati- num electrode and surprisingly, yielded a higher electrochemical response. The best defined anodic peak was obtained with 0.2 mol L-1 KNO3 pH 5.0 after 25 min of accumulation time. Using these conditions, the cali- bration plot was linear over the range 1 × 10?8 to 5 × 10?6 mol L-1 Cd2+. The precision was examined by car- rying out eight replicate measurements at a concentration of 2.5 ×10?5 mol L-1;the coefficient of variation was 2.9%. The method was applied to the determination of the analyte in river water samples. The interfere- ence of other metal ions on the voltammetric response of Cd(II) was studied. The HAP films was clearly ob- served in the SEM images and characterized by X-ray diffraction, IR spectroscopy and chemical analysis.