首页> 中文期刊> 《美国分析化学(英文)》 >Determination of Palladium II in 5 Pd/BaSOsub4/subby ICP-MS with Microwave Digestion, and UV-VIS Spectrophotometer

Determination of Palladium II in 5 Pd/BaSOsub4/subby ICP-MS with Microwave Digestion, and UV-VIS Spectrophotometer

         

摘要

Determination palladiums have been reported 5% (w/w) Pd/BaSO4 known as Rosenmund Catalyst. The determination of palladium II known as Rosenmund Catalyst is always an expensive procedure usually involving procedures such as flame atomic absorption spectrophotometry, emission spectrometry, and many spectrophotometric methods. In this study, palladium II in 5% Pd/BaSO4, was synthesized and employed to develop an extractive UV-Visible Spectrophotometric, and an inductively coupled plasma mass spectrometry ICP/MS methods for the determination of palladium II. Specification for Pd is 4.85% to 5.10%;the result was 4.97% for the UV-Visible spectrophotometer and 4.90% for the ICP/MS. Both results meet the requirements.