A simple method to objectively and simultaneously measure eye's longitudinal and transverse chromatic aberrations was proposed.A dual-wavelength wavefront measurement system using two Hartmann-Shack wavefront sensors was developed.The wavefronts of the red(639.1 nm)and near-infrared(786.0 nm)lights were measured simultaneously for different positions in the model eye.The chromatic wavefronts were converted into Zernike polynomials.The Zernike tilt cofficient(irst term)was used to calculate the transverse chromatic aberration along the ax-direction,while the Zernike defocus coefficient(fourth term)was used to calculate the longi-tudinal chromatic aberration.The measurement and simulation data were consistent.