首页>
中文期刊>中国中西医结合杂志:英文版
>Charge Couple Device-Based Systemfor3-di mensional Real Ti me Positioning on the Assessment of Segmental Range of Motion of Lumbar Spine
Charge Couple Device-Based Systemfor3-di mensional Real Ti me Positioning on the Assessment of Segmental Range of Motion of Lumbar Spine
Objective: To observe the tested results of the segmental range of motion (ROM) of lumbar spine by charge couple device (CCD)-based system for 3-dimensional real-time positioning (CCD system), and to analyze its clinical significance.Methods: Seven patients with lumbar joint dysfunction and 8 healthy subjects were tested twice by the CCD-based system with an interval of 10 min. Results: The ROM of the patients was obviously lesser than that of the healthy subjects. The measuring data of segmental ROM of lumbar spine by CCD system is correlated significantly to the same data checked later on the same subjects in every direction of the movements. The differences between two checks are usually less than 1 degree. Conclusion: Specially designed CCD based system for 3-dimensional real-time positioning could objectively reflect the segmental ROM of lumbar spine. The system would be of great clinical significance in the assessment of the biomechanical dysfunction of lumbar spine and the effect of the treatment applied.
展开▼