The Analysis of Effective Radiation Dose during PET/CT Examination

         

摘要

Objective: In order to reduce the effective dose level of the patients to explore the best scanning conditions, various factors affecting the level of PET/CT examination were analyzed. Methods:The effective doses of CT and pet were calculated respectively during PET/CT examination, Scanning parameters of CT and PET were optimized to reduce the effective dose of patients to a minimum while maintaining the image quality. Results:The tube voltage and tube current of CT, the injection dose during PET examination can affect the effective dose of the patients significantly. Conclusion:According to different types of patients and diseases to develop personalized scanning program, the effective dose level of the patients can be significantly reduced.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有