SL-IC放射治疗模拟机高压电缆击穿原因分析

         

摘要

目的:解决我科放射治疗模拟机高压电缆几个月就被击穿的现象;方法:通过对SL-IC放射治疗模拟机高压电路的分析,总结引起该机高压电缆易被击穿的原因;结果:在最终选用厂家提供的合格的高压发生器后使故障解决;结论:放射治疗模拟机高压电路故障隐蔽,要和厂家密切联系,并经过反复试验和认真总结,快速完成机器维护工作,更好地服务临床.%Objective: To solve the problem of breakdown in high voltage eleclrocircuit of radiotherapy simulator in a couple of months in our department. Methods The high voltage electrocircuit of radiotherapy simulator was analyzed to determine reasons for its breakdown. Results: After selecting an eligible high voltage generator provided by manufacturer, the failure was solved finally. Conclusion: As for solution of the failure about the high voltage electrocircuit of radiotherapy simulator, we must keep in touch with the manufacturer to rapidly perform maintenance to serve the clinic better.