首页> 中文期刊> 《化学分析计量》 >原子吸收光谱法测定纯铼中痕量元素钠

原子吸收光谱法测定纯铼中痕量元素钠

         

摘要

采用火焰原子吸收光谱法(AAS)法测定纯铼中痕量元素钠。对仪器工作条件如光谱通带、灯电流、乙炔流量进行了分析,优化了原子吸收测量条件。Na的质量分数在0.0001%~0.001%范围内与吸光度A线性关系良好,线性方程为A=176.1c+0.004,相关系数r=0.9997,方法检出限为0.0014μg/mL。钠测量结果的相对标准偏差不大于7.29%(n=8),回收率为93.3%~107.4%。该法操作简便、分析时间短、干扰少。%The determination of trace Na in pure rhenium was investigated by AAS,measuring conditions such as light spectrum pass band,lamp current,air-C2H2 flow rate were analyzed,and the measuring conditions of AAS was optimized. The mass fraction of Na was linear with the absorbance in the range of 0.000 1%-0.001%,the regression equation was A=176.1c+0.004,the correlative coefficient r=0.999 7. The detection limit was 0.001 4μg/mL. The relative standard deviations of determination results were less than 7.29%(n=8),the recovery was 93.3%-107.4%. The method is simple, short analysis time, and less interference.

著录项

  • 来源
    《化学分析计量》 |2013年第6期|29-31|共3页
  • 作者

    任慧; 孙建新;

  • 作者单位

    航空材料检测与评价北京市重点实验室;

    中航工业北京航空材料研究院;

    中航 试金石 检测科技有限公司;

    北京 100095;

    航空材料检测与评价北京市重点实验室;

    中航工业北京航空材料研究院;

    中航 试金石 检测科技有限公司;

    北京 100095;

  • 原文格式 PDF
  • 正文语种 chi
  • 中图分类 光化学分析法(光谱分析法);
  • 关键词

    原子吸收光谱; 铼; 钠;