首页> 外文期刊>高能物理与核物理计算国际会议公报:英文版 >Calculation of Energy Response of Cylindrical G—M Tubes with EGS4 Monte Carlo Code
【24h】

Calculation of Energy Response of Cylindrical G—M Tubes with EGS4 Monte Carlo Code

机译:用EGS4蒙特卡洛法计算圆柱形GM管的能量响应

获取原文
获取原文并翻译 | 示例
       

摘要

The energy responses of Ka of two types of cylindrical G-M counter tubes were calculated using an electron-photon cascade Monte Carlo code,EGS4.One type of the G-M counter tubes was GJ4401 (sensitive length 9cm,diameter 1cm),the other was J5 (sensitive length 2cm,diameter 0.3cm),the restricted sampling technique of source photon was used.Good tendency agreements between the simulations and experiments were achieved for gamma radiation with energies ranging from 40keV to 1.25 MeV.For GJ4401,the difference of response between simulations and experiments at 662 keV was 34% and for J5 the difference was 27%.
机译:使用电子-光子级联蒙特卡罗代码EGS4计算两种圆柱形GM计数管的Ka的能量响应.GM计数管的一种类型为GJ4401(敏感长度9cm,直径1cm),另一种为J5(敏感长度2厘米,直径0.3厘米),使用了源光子的受限采样技术。对于能量范围从40keV到1.25 MeV的伽马辐射,仿真与实验之间取得了良好的趋势一致性。对于GJ4401,仿真之间的响应差异而662 keV的实验为34%,J5的差异为27%。