Establishment and Evaluation of a Method for Analyzing Atmospheric Volatile Organic Compounds

机译:大气挥发性有机物分析方法的建立与评价

获取原文
获取原文并翻译 | 示例
       

摘要

An automated cumulative sampling system and a method that combines a two-step cryo-concentrated system and gas chromatography/mass spectrometry (CCS-GC/MS) axe introduced. The method is evaluated by a set of special experiments and the results are presented. The lowest measurement detection limit was expanded from 10-6 nmol mol-1 to 10-12 nmol mol-1 by using CCS-GC/MS instead of the simpler method of gas chromatography/mass spectrometry (GC/MS), with the average responsible factor of 39 object compounds being 2.9×10-12. When the volume of air sample reached 1000 cm3, the lowest detection limit reached up to 7×10-12-40×10-12 nmol mol-1. The CCS-GC/MS method can potentially identify all objective chemical species in an atmospheric sample, with an average 2.5 s bias error of retention time for 39 gas chromatography (GC) peaks. Within the range 0-400×10-9 nmol mol-1, the concentration of 39 kinds of objective compounds can be individually calculated very accurately by a standard curve [average r2(coefficient of determination) value of above 0.99]. The recovery efficiency was 88%-111%, with an average of 100.8% +5.6%. The bias error of precision was 2%-14%, with an average of 6.6%.
机译:介绍了一种自动累积采样系统和结合了两步低温浓缩系统和气相色谱/质谱法(CCS-GC / MS)的方法。通过一组特殊实验对方法进行了评估,并给出了结果。通过使用CCS-GC / MS代替最简单的气相色谱/质谱法(GC / MS),将最低检测限从10-6 nmol mol-1扩展到10-12 nmol mol-1 39种目标化合物的有效因子为2.9×10-12。当空气样品的体积达到1000 cm3时,最低检测限达到7×10-12-40×10-12 nmol mol-1。 CCS-GC / MS方法可以潜在地识别大气样品中的所有目标化学物质,其中39个气相色谱(GC)峰的保留时间平均偏差为2.5 s。在0-400×10-9 nmol mol-1的范围内,可以通过标准曲线[r2(测定系数)的平均值大于0.99)非常精确地分别计算39种目标化合物的浓度。回收效率为88%-111%,平均为100.8%+ 5.6%。精度偏差为2%-14%,平均为6.6%。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号