首页> 外文学位 >Comparison of video and CCD cameras in online portal imagers calibrated for dosimetry.
【24h】

Comparison of video and CCD cameras in online portal imagers calibrated for dosimetry.

机译:校准了剂量学的在线门户成像仪中的摄像机和CCD摄像机的比较。

获取原文
获取原文并翻译 | 示例

摘要

Video-based electronic portal imaging devices (EPIDs) are the most common type of portal imager found in clinical practice. In addition to providing a method to detect and correct patient set-up errors and record treatments, possibilities exist for the dosimetric use of video-based EPIDs. To determine if a particular system is suitable for dosimetry, the entire system and especially the camera should be thoroughly tested and calibrated. Comparisons were made of the performance of a Newvicon camera and a CCD camera in a video-based EPID. Tests were performed on the cameras to investigate linearity, noise, frame grabber effects, clamping error, and lens vignetting. Tests of the entire imaging system were performed to determine the behaviour of the veiling glare, the uniformity of the spatial sensitivity, and the EPID response with changes in field size and phantom thickness. The veiling glare was caused by several factors: the mirror, the cameras themselves, and the phantom scatter and beam hardening. The magnitude of the glare was spatially dependent. The EPID response was not spatially uniform, but increased by approximately 25% in the region farthest from the screen-mirror junction. The cause of this increase was determined to be the mirror. The dose-corrected EPID response with changes in phantom thickness was nearly flat for medium-sized fields, although it did increase slightly for large fields and decrease slightly for small fields. The overall performance of the CCD camera was found to be significantly better than that of the Newvicon camera. Central axis dosimetry would be straightforward to calibrate; however, the spatially dependent glare observed in the system would make it extremely difficult to deconvolve an EPID image to provide a two-dimensional dose map.
机译:基于视频的电子门户成像设备(EPID)是临床实践中最常见的门户成像器。除了提供一种检测和纠正患者设置错误并记录治疗的方法外,还存在用于基于视频的EPID的剂量学使用的可能性。为了确定特定系统是否适合剂量测定,应对整个系统(尤其是摄像机)进行彻底的测试和校准。对基于视频的EPID中的Newvicon相机和CCD相机的性能进行了比较。在相机上进行了测试,以调查线性度,噪声,图像捕捉器效果,夹持误差和镜头渐晕。进行了整个成像系统的测试,以确定遮盖眩光的行为,空间灵敏度的均匀性以及随着视场大小和幻像厚度的变化而产生的EPID响应。蒙面的眩光是由多种因素造成的:镜子,相机本身以及幻像散射和光束硬化。眩光的大小与空间有关。 EPID响应在空间上不均匀,但在离屏幕镜接合点最远的区域中增加了约25%。确定这种增加的原因是镜子。剂量校正的EPID响应随着幻像厚度的变化,对于中型场几乎是平坦的,尽管对于大场而言确实有所增加,而对于小场而言确实有所下降。发现CCD相机的整体性能明显优于Newvicon相机。中心轴剂量法将易于校准;然而,在系统中观察到的空间依赖性眩光将使对EPID图像进行反卷积以提供二维剂量图极为困难。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号