A limited evaluation of Ultraviolet Multifilter Rotating Shadowband Radiometer (UVMFRSR) based monitoring was conducted for instruments operated at the Mauna Loa Observatory, Hawaii 2/1/04 - 5/31/04. Two UV radiometers were installed, an UVMFRSR (shadowband) and a tracker mounted UVMFR (no shadowband). Measurements from the two UVMFR's, collocated Brewer and Dobson UV spectrophotometers, and the TOMS satellite, allowed a number of intercomparisons. The Langley-Forgan calibration technique proved more effective than lamp or Langley-SLR (simple linear regression) techniques when applied to an UVMFR that exhibited substantial change in response. Direct irradiance from an UVMFRSR compared very well with direct irradiance measured by a tracker mounted UVMFR for full-sun days (r~2's = 1). Mean ratios of the UV-MFRSR column ozone retrievals to the collocated Dobson and Brewer were 1.015 and 1.011 respectively. The mean ratio of UVMFRSR ozone to TOMS satellite ozone was 0.907. The UVMFRSR advantages of relatively low cost, unattended operation, automated calibration stability checks using Langley plots, and minimal maintenance make it a unique instrument for column ozone measurement.
机译:对在夏威夷的莫纳罗亚天文台,2004年2月1日至2004年5月31日运行的仪器进行了基于紫外线的多滤光片旋转暗带辐射计(UVMFRSR)监测的有限评估。安装了两个紫外线辐射计,一个UVMFRSR(阴影带)和一个安装有跟踪器的UVMFR(无阴影带)。来自两个并置的Brewer和Dobson紫外分光光度计以及TOMS卫星的UVMFR的测量结果进行了许多比较。当将Langley-Forgan校准技术应用于表现出实质性变化的UVMFR时,它比lamp或Langley-SLR(简单线性回归)技术更有效。来自UVMFRSR的直接辐照度与安装在跟踪器上的UVMFR在全日照下测得的直接辐照度非常好(r〜2's = 1)。 UV-MFRSR色谱柱臭氧回收量与并置的Dobson和Brewer的平均比分别为1.015和1.011。 UVMFRSR臭氧与TOMS卫星臭氧的平均比值为0.907。 UVMFRSR具有相对较低的成本,无人值守的操作,使用Langley图进行自动校准稳定性检查以及维护最少的优点,使其成为用于色谱柱臭氧测量的独特仪器。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号