首页> 外文会议>Society for Biomaterials Transaction Annual Meeting vol.29 pt.1 >MULTI-LAB EVALUATION OF HIP SIMULATOR TEST CONDITIONS ON RELATIVE WEAR RATES OF CROSSLINKED POLYETHYLENES
【24h】

MULTI-LAB EVALUATION OF HIP SIMULATOR TEST CONDITIONS ON RELATIVE WEAR RATES OF CROSSLINKED POLYETHYLENES

机译:基于交联聚乙烯相对磨损率的HIP仿真器测试条件的多实验室评估

获取原文
获取外文期刊封面目录资料

摘要

Hip wear simulations are an effective method to assess the clinical potential of crosslinked relative to conventional UHMVVPE materials. In general, recently published short and mid term clinical studies for crosslinked and conventional UHMWPE indicate good agreement with simulation results. '' However, there remain substantial differences in the preferred test conditions among laboratories internationally. This study investigated the effect of test conditions from 3 test centers on the wear of conventional and crosslinked poly ethylenes. To permit comparisons, each center included common UHMWPE materials, and multiple serum protein concentrations. Conditions unique to each test center, including load magnitude and profile, simulator kinematics and serum chamber volume were unaltered from their earlier studies.
机译:髋关节磨损模拟是评估相对于常规UHMVVPE材料的交联临床潜力的有效方法。总的来说,最近发表的关于交联的和常规的UHMWPE的短期和中期临床研究表明与模拟结果有很好的一致性。 ''但是,国际实验室在首选测试条件方面仍然存在实质性差异。这项研究调查了3个测试中心的测试条件对常规聚乙烯和交联聚乙烯的磨损的影响。为了进行比较,每个中心都包括通用的UHMWPE材料和多种血清蛋白浓度。每个测试中心特有的条件(包括负载大小和轮廓,模拟器运动学和血清腔体积)在其早期研究中均未改变。