DISCHARGE COEFFICIENT CHANGES IN CRITICAL FLOW VENTURI NOZZLES AFTER SEVERE FIELD SERVICE

机译:严重现场维修后,临界流量文丘里喷嘴的出水系数变化

获取原文
获取原文并翻译 | 示例

摘要

This paper presents calibration accuracy results of fivernsonic nozzles after three hundred hours of field service. Inrn1995, the sonic nozzles were removed from the system forrninspection. Upon removing the nozzles, a thin layer ofrnelemental sulfur, compressor oil, and glycol was found on thernsurface of the nozzles. The nozzles were then re-calibrated "asrnfound" at the Colorado Engineering Experiment Station, Inc.,rn(CEESI) to determine any drift in the discharge coefficient duernto the surface contamination. K-Lab in Norway also foundrnsulfur deposits on their nozzle system in 1989 and laterrninstalled zinc oxide scrubber to remove low level H_2S in therngas stream.rnSonic nozzles were chosen as on-site transfer provers tornverify and calibrate natural gas custody transfer turbine metersrnin Chevron’s Venice flow metering station. The facility,rnlocated in Venice, Louisiana, USA, is owned and operated byrnChevron U.S.A., Inc., since 1987 to meter dry processedrnnatural gas downstream of a gas plant. The sales capacity isrn14.16 × 10~6 std m3 (500 MMSCFD) at a nominal operatingrnpressure of 6.9 MPa (1,000 psig). The system includes a bankrnof five sonic nozzles and a master turbine meter, all connectedrnin series with three 30.48-cm (12-in.) sales turbine meters.rnThe sales meters were tested monthly against the master meterrnfor accuracy. The accuracy of the master meter was thenrnverified with sonic nozzles every three months. If any turbinernmeters did not meet the accuracy requirement, on-siterncalibration was allowed using sonic nozzles to develop a newrnk-factor for the meter. The sonic nozzles were constructed,rncalibrated, and operated under the guidelines of ANSI/ASMErnMFC-7M-1987 standard.rnThe "as found" discharge coefficient of all fiverncontaminated nozzles were in the range of 0.1% to 0.3% lowerrnthan the original calibration data. However, the shift was stillrnwithin the measurement uncertainty of the CEESI calibrationrnfacility. The results clearly indicated that a sonic nozzle is arnrobust flow proving device and that measurement accuracyrnwas affected slightly when operating in a hostile environment.
机译:本文介绍了经过300小时的现场服务后,firnsonic喷嘴的校准精度结果。 1995年,将声波喷嘴从系统中移除,进行检查。卸下喷嘴后,在喷嘴的表面发现了一层薄薄的元素硫,压缩机油和乙二醇。然后在科罗拉多工程实验站有限公司(CEESI)对喷嘴进行“校准”,以确定由于表面污染导致的排放系数的任何漂移。挪威的K-Lab于1989年还在喷嘴系统上发现了硫沉积物,后来安装了氧化锌洗涤塔以去除气流中的低水平H_2S。rn声波喷嘴被选为现场传输证明,以验证和校准天然气贸易交接透平涡轮流量计在雪佛龙公司威尼斯流程中的应用计量站。该设施位于美国路易斯安那州的威尼斯,自1987年以来一直由雪佛龙美国公司拥有和运营,以计量天然气厂下游的干燥天然气。在6.9 MPa(1,000 psig)的额定工作压力下,销售能力为14.16×10〜6 std m3(500 MMSCFD)。该系统包括一个带有五个超声波喷嘴的bankrnof和一个主涡轮流量计,所有涡轮流量计均与三个系列的30.48厘米(12英寸)销售涡轮流量计相连。rn每月对销售仪表的主准确性进行测试。然后每三个月用声波喷嘴检查一次主仪表的精度。如果任何涡轮流量计不满足精度要求,则可以使用声波喷嘴进行即时校准,以开发出新的流量计因子。声波喷嘴是按照ANSI / ASMErnMFC-7M-1987标准进行构造,校准和运行的。所有五个被污染的喷嘴的“发现”排放系数均比原始校准数据低0.1%至0.3%。但是,这种变化仍在CEESI校准设施的测量不确定性范围内。结果清楚地表明,声波喷嘴是可靠的流量证明装置,并且在恶劣的环境中工作时,测量精度会受到轻微影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号