Characterization of Light Distribution and Optimization of Detector Position for Multiple Reference Optical Coherence Tomography

机译:多参考光学相干层析成像的光分布特征和探测器位置的优化

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A time-domain optical coherence tomography system with respect to multiple reflections in the reference arm is simulated with Zemax and the optimal detector position is discussed based on the wavefront error. Multiple reference optical coherence tomography (MR-OCT) is a novel optical imaging platform using a single miniature actuator and partial mirror to recirculate light multiple times within the interferometric system. The recirculation corresponds to multiple path delays extending effectively the axial scan-range. The simulation of MR-OCT with lenses of 5 mm diameter shows that it is suitable to build a compact system using low-cost components. Subsequently MR-OCT enables an affordable miniature OCT for low recourse settings and personal or point-of-care applications.
机译:利用Zemax模拟了相对于参考臂中多次反射的时域光学相干层析成像系统,并基于波前误差讨论了最佳探测器位置。多参考光学相干断层扫描(MR-OCT)是一种新颖的光学成像平台,使用单个微型致动器和部分反射镜在干涉测量系统中多次循环光。再循环对应于有效地扩展轴向扫描范围的多路径延迟。用直径为5 mm的透镜对MR-OCT进行的仿真表明,适合使用低成本组件构建紧凑的系统。随后,MR-OCT实现了价格低廉的微型OCT,适用于低资源设置和个人或即时护理应用。