首页> 外文会议>Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE >Time resolution performance of an electro-optical-coupled PET detector for time-of-flight PET/MRI
【24h】

Time resolution performance of an electro-optical-coupled PET detector for time-of-flight PET/MRI

机译:用于飞行时间PET / MRI的电光耦合PET检测器的时间分辨率性能

获取原文

摘要

Combining PET with MRI in a single system provides clinicians with complementary anatomical and molecular information. However, current combined PET/MRI systems do not have time-of-flight (ToF)-PET capabilities due to certain design choices. We have developed an MRI-compatible front-end electronic system with fast timing PET capabilities. Our approach employs, a fast arrival time pickoff comparator to digitize the timing information, and a laser diode to drive a fiber-optic cable to optically transmit timing information. The comparator and electo-optical link show 17ps +/− 0.2ps fwhm jitter in response to a fast digital pulse. When coupled with LSO scintillation crystals and Hamamatsu MPPC silicon photo-multipliers (SiPM) the comparator and electro-optical link achieved a time resolution of 265ps +/− 6.3ps with 3×3×20mm crystals and 167 +/− 3.6ps fwhm with 3×3×5mm crystals.
机译:在单个系统中将PET与MRI相结合,可为临床医生提供互补的解剖和分子信息。但是,由于某些设计选择,当前的组合PET / MRI系统不具备飞行时间(ToF)-PET功能。我们已经开发了具有MRI兼容功能的前端电子系统,并具有快速定时PET功能。我们的方法采用快速到达时间传感器比较器将时序信息数字化,并使用激光二极管驱动光缆来光学传输时序信息。比较器和电光链路响应快速数字脉冲显示17ps +/- 0.2ps fwhm抖动。当与LSO闪烁晶体和Hamamatsu MPPC硅光电倍增器(SiPM)结合使用时,比较器和电光链路的3×3×20mm晶体和167 +/- 3.6ps fwhm的时间分辨率达到265ps +/- 6.3ps。 3×3×5mm晶体。