Concept for Tremor Compensation for a Handheld OCT-Laryngoscope

机译:手持式OCT喉镜震颤补偿的概念

获取原文
获取原文并翻译 | 示例

摘要

Optical coherence tomography (OCT) is a non-invasive imaging technique which can create optical tissue sections, enabling diagnosis of vocal cord tissue. To take full advantage from the non-contact imaging technique, OCT was adapted to an indirect laryngoscope to work on awake patients. Using OCT in a handheld diagnostic device the challenges of rapid working distance adjustment and tracking of axial motion arise. The optical focus of the endoscopic sample arm and the reference-arm length can be adjusted in a range of 40 mm to 90 mm. Automatic working distance adjustment is based on image analysis of OCT B-scans which identifies off depth images as well as position errors. The movable focal plane and reference plane are used to adjust working distance to match the sample depth and stabilise the sample in the desired axial position of the OCT scans. The autofocus adjusts the working distance within maximum 2.7 seconds for the maximum initial displacement of 40 mm. The amplitude of hand tremor during 60 s handheld scanning was reduced to 50 % and it was shown that the image stabilisation keeps the position error below 0.5 mm. Fast automatic working distance adjustment is crucial to minimise the duration of the diagnostic procedure. The image stabilisation compensates relative axial movements during handheld scanning.
机译:光学相干断层扫描(OCT)是一种非侵入性成像技术,可以创建光学组织切片,从而能够诊断声带组织。为了充分利用非接触式成像技术的优势,OCT适用于间接喉镜,用于清醒患者。在手持式诊断设备中使用OCT带来了快速工作距离调整和轴向运动跟踪的挑战。内窥镜样品臂的光学焦点和参考臂的长度可以在40 mm至90 mm的范围内调节。自动工作距离调整基于OCT B扫描的图像分析,该图像可识别深度图像以及位置误差。可移动焦平面和参考平面用于调整工作距离以匹配样品深度,并使样品稳定在OCT扫描的所需轴向位置。自动对焦可在最多2.7秒内调整工作距离,最大初始位移为40 mm。手持扫描60 s期间的手震幅度降低到50%,这表明图像稳定可将位置误差保持在0.5 mm以下。快速的自动工作距离调节对于最小化诊断过程的持续时间至关重要。图像稳定功能可补偿手持式扫描过程中的相对轴向运动。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号