Lockheed Martin's Sunnyvale facility operates 4 large thermal vacuum chambers, each having a data channel count in excess of 700 channels. Recently, the data acquisition systems attached to these chambers were upgraded. The new system was expected to meet an accuracy specification of +- 1.5 degree C over a temperature range of 200 to - 200 degree C. The initial installation failed to meet this specification for roughly 20 percent of its data channels. This paper discusses the sources of error that caused the system to fail calibration and how they were addressed.
机译:洛克希德·马丁公司位于桑尼维尔的工厂运营着4个大型热真空室,每个真空室的数据通道数超过700个。最近,连接到这些腔室的数据采集系统得到了升级。预计新系统将在200至200摄氏度的温度范围内满足±1.5摄氏度的精度规格。最初的安装未能满足其大约20%数据通道的规格要求。本文讨论了导致系统无法校准的错误源以及如何解决这些错误。