首页> 外文会议>International Conference on Biomedical Engineering and Informatics >Noninvasive blood pressure analysis using empirical mode decomposition (EMD) - Based oscillometric method
【24h】

Noninvasive blood pressure analysis using empirical mode decomposition (EMD) - Based oscillometric method

机译:基于实证分解的非血压血压分析(EMD)的示波法

获取原文

摘要

This study proposes a new method to extract oscillometric pulses for noninvasive blood pressure measurement using empirical mode decomposition, based on the oscillometric method. The mean and standard deviation of the present method in systolic and diastolic pressure are 1.6 and 5.2, and 0.7 and 5.2, respectively, which not only passes the AAMI (Association for the Advancement of Medical Instrumentation) and EN (European Norm) blood pressure measurement criteria, but also performs more accurate measurements than commercial noninvasive blood pressure measurement devices.
机译:本研究提出了一种利用经验模式分解提取用于非血压血压测量的示波器脉冲的新方法,基于示波法。本发明方法在收缩压和舒张压中的平均值和标准偏差分别为1.6和5.2,以及0.7和5.2,不仅通过AAMI(医疗仪器的进步协会)和EN(欧洲常态)血压测量标准,但也比商业非侵入性血压测量装置执行更准确的测量。