Dosimetric Modeling of Mammography Using the Monte Carlo Code PENELOPE and Its Validation

机译:蒙特卡罗代码普莱洛普乳房X线照相X线X线X线X线及其验证

获取原文

摘要

The Monte Carlo code PENELOPE (version 2014) and its software package PenEasy have been employed to closely simulate a commercial mammography system: Selenia Dimensions system (Hologic, Bedford, Mass) for dosimetric study. Normalized glandular dose (D_gN) was derived for specific filter combinations including tungsten/aluminum (W/Al), tungsten/rhodium (W/Rh) and tungsten/silver (W/Ag) for a range of clinically used voltages (kV). Breast models with thickness 2 cm, 4.5 cm and 8 cm and respective glandularity 100 %, 50 % and 0.1 % were included in the simulations. Derived D_gN values were in good agreement with well-accepted published values, with mean differences of 1.30 %, 2.17 % and 0.17 % for the W/Al, W/Rh and W/Ag anode/filter combinations, respectively. In the next phase of the study, this validated model will be modified to perform patient-specific dosimetry with an improved breast model and more realistic simulation of commercial mammography systems.
机译:蒙特卡罗代码Penelope(2014年版)及其软件包Peneasy被用来密切模拟商业乳房系统:Selenia尺寸系统(Hologic,Bedford,Mass)进行剂量测定研究。归一化腺剂量(D_GN)衍生出特定的过滤器组合,包括钨/铝(W / Al),钨/铑(W / RH)和钨/银(W / AL),用于一系列临床用电压(KV)。厚度为2厘米,4.5厘米和8cm,腺体的乳房模型100%,50%和0.1%均包括在模拟中。衍生的D_GN值与接受良好的公布值良好,对于W / Al,W / RH和W / AG阳极/滤波器组合的平均差异为1.30%,2.17%和0.17%。在该研究的下一阶段,将被修改该验证的模型,以通过改进的乳房模型和商业乳房X线摄影系统的更现实模拟来进行患者特定的剂量测定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号