Removing power line interference from ECG signal using adaptive filter and notch filter

机译:使用自适应滤波器和陷波滤波器从ECG信号中取出电源线干扰

获取原文

摘要

Performance of two adaptive filters, such as, normalized least-mean-square (NLMS) adaptive filter and recursive-least-square (RLS) adaptive filter are compared with a traditional notch filter both in time and frequency domains to remove the power line interference from the ECG signal. The power spectral density (PSD) and spectrogram analyses are also performed. Different performance parameters, such as, SNR, %PRD and MSE are also calculated. Real time recorded data from the benchmark MIT-BIH arrhythmia database has been used. The result demonstrates superior performance of adaptive NLMS filter for removing power line interference over adaptive RLS and notch filters.
机译:两个自适应滤波器的性能,例如归一化最小平均方形(NLMS)自适应滤波器和递归 - 最小二乘(RLS)自适应滤波器与在时间和频域中的传统凹口过滤器进行比较,以去除电力线干扰来自ECG信号。还执行功率谱密度(PSD)和频谱图分析。还计算不同的性能参数,例如SNR,%PRD和MSE。已经使用了从基准MIT-BIH心律失常数据库中记录的实时记录数据。结果表明,自适应NLMS滤波器的卓越性能,用于去除自适应RLS和缺口滤波器的电力线路干扰。