首页> 外文会议>Nuclear Science Symposium >Results of L3 BGO calorimeter calibration using an RFQ accelerator
【24h】

Results of L3 BGO calorimeter calibration using an RFQ accelerator

机译:使用RFQ加速器的L3 BGO热计校准结果

获取原文

摘要

A novel calibration system based on a radio-frequency-quadrupole (RFQ) accelerator has been installed in the L3 experiment. Radiative capture of 1.85 MeV protons from the RFQ accelerator in a lithium target produces a high-intensity flux of 17.6 MeV photons which are used to calibrate 11,000 crystals of the L3 BGO calorimeter. In this paper, we present results of the first fully successful RFQ run, taken in November 1997. We achieved a calibration precision of 0.6% in the barrel and 0.7% in the endcaps of the L3 BGO calorimeter. The RFQ-97 calibration proved to be the best one of all calibration techniques used in the L3 experiment.
机译:L3实验安装了基于射频 - 四极(RFQ)加速器的新型校准系统。从锂靶中的RFQ加速器辐射捕获1.85meV质子的辐射捕获产生的高强度通量为17.6meV光子,用于校准L3 BGO量热计的11,000个晶体。在本文中,我们提出了第一个完全成功的RFQ运行的结果,于1997年11月拍摄。我们在L3 BGO热量计的端盖中实现了0.6%的校准精度和0.7%。 RFQ-97校准被证明是L3实验中使用的所有校准技术中最好的一种。